4. Junior User Guidebook
Overcompensated Compensated correctly Under compensated
3. Repeat
Fig. 4-7 Displayed Waveforms of the Probe Compensation
the steps mentioned if needed.
Fig. 4-8 Adjust Probe
How to Set the Probe Attenuation Coefficient
The probe has several attenuation coefficients, which will influence the vertical scale
e attenuation coefficient in the menu of oscilloscope:
H2
tton to display the Probe menu; select the proper value corresponding
lid all the time before it is changed again.
factor of the oscilloscope.
To change or check the prob
(1) Press the function menu button of the used channels (CH1 MENU or C
MENU).
(2) Press H3 bu
to the probe.
This setting will be va
Caution: The default attenuation coefficient of the probe on the instrument is
he set value of the attenuation switch in the probe is
preset to 10X.
Make sure that t
the same as the menu selection of the probe attenuation coefficient in
the oscilloscope.
The set values of the probe switch are 1X and 10X (see Fig. 4-9).
14