8. Technical Specifications
8. Technical Specifications
Unless otherwise specified, the technical specifications applied are for SDS5032E(V)
only, and Probes attenuation set as 10X. Only if the oscilloscope fulfills the following
two conditions at first, these specification standards can be reached.
This instrument should run for at least 30 minutes continuously under the
specified operating temperature.
If change of the operating temperature is up to or exceeds 5 , ℃ do
"Self-calibration" procedure (see "How to Implement Self-calibration" on P15).
All specification standards can be fulfilled, except one(s) marked with the word
"Typical".
Performance Characteristics Instruction
Bandwidth
30MHz
Channel
2 + 1 (External)
Mode Normal, Peak detect, Averaging
Dual CH Up to 125MS/s
Acquisition
Sample rate
(real time)
Single CH Up to 250MS/s
Input coupling DC, AC , Ground
Input impedance 1MΩ±2%, in parallel with 10pF±5pF
Probe attenuation
factor
1X, 10X, 100X, 1000X
Max. input voltage 400V (PK-PK) (DC + AC PK-PK)
Channel –channel
isolation
50Hz: 100 : 1
10MHz: 40 : 1
Input
Time delay between
channel(typical)
150ps
Dual CH 5S/s~125MS/s
Sampling rate range
Single CH 5S/s~250MS/s
Interpolation (sin x)/x
Dual CH
Max Record length
Single CH
≤Max sampling rate
10K
Scanning speed
(S/div)
4ns/div~100s/div,
step by 1~2~4
Horizontal
System
Sampling rate /
relay time accuracy
±100ppm
81