4. Junior User Guidebook
(1) Press the function menu button of the used channels (CH1 MENU or CH2
MENU).
(2) Press H3 button to display the Probe menu; select the proper value corresponding
to the probe.
This setting will be valid all the time before it is changed again.
Note: The attenuation coefficient of the probe in the menu is preset to 10X
when the oscilloscope is delivered from the factory.
Make sure that the set value of the attenuation switch in the probe is
the same as the menu selection of the probe attenuation coefficient in
the oscilloscope.
The set values of the probe switch are 1X and 10X (see Fig. 4-10).
Fig.4-10 Attenuation Switch
Note: When the attenuation switch is set to 1X, the probe will limit the
bandwidth of the oscilloscope in 5MHz. If it is needed to use the full
bandwidth of the oscilloscope, the switch must be set to 10X.
How to Use the Probe Safely
The safety guard ring around the probe body protects your finger against the electric
shock, shown as Fig. 4-11.
Fig. 4-11 Finger Guard
User Manual of Smart Digital Storage Oscilloscope
15