Page 48
© 2024 PacBio. All rights reserved. For research use only. Not for use in diagnostic procedures.
PN 103-140-400 REV05 JUN2024
9.3
Close both workspaces and select “Confirm”.
9.4
Select Start” to begin the system check tests.
9.5
Once tests complete, results can be found in D:\PacBio\Metrics\PreRun
9.6
When finished, select “Close” to go back to the
Home
screen.
Peri pump prelim check test
9.7
Select “Open workspaces”. Load an empty waste bottle and wash pack filled with 1L of instrument wash
reagent into the system.
9.8
Load the wash cartridges onto the system and check the engagement by gently pulling upward on both
cartridges to verify they do not lift out of the system.
9.9
Close both workspaces and select “Confirm”.
9.10
Select Start” to begin the system check test.
9.11
Once the test is complete, results can be found in D:\PacBio\Metrics\PreRun
9.12
When finished, select “Close” to go back to the
Home
screen.