Parker Hannifin S.p.A.  - S.B.C. Division                                               HPD N User’s Manual 
 
37
4.5. Calibrating speed loop 
 
IMPORTANT CONCEPTS 
 
SPEED LOOP: the main task of any drive is to control motor speed so that it follows, as 
faithfully as possible, the required speed, which is generally referred to as the REFERENCE. 
High fidelity reproduction of the reference implies that motor speed is identical to reference 
speed in static conditions and also that it follows speed requirements closely also in the case 
of sudden changes in the speed reference (dynamic conditions). To perform this task the drive 
must be informed of various characteristics of the motor and the mechanical system to which 
the motor is coupled; this type of information is set in the CALIBRATION PARAMETERS. 
ERROR: the error refers to the difference between reference speed and motor speed. The 
error value is utilised by the speed control loop to calculate (using calibration parameters) 
how much current to supply to the motor. 
TORQUE: current flowing through the motor windings is converted into torque which 
allows the motor to accelerate and decelerate. 
GAIN: in consideration of the typical applications of the HPD drive, the expression “gain” 
in this section of the manual refers to “stiffness”. To understand the concept of stiffness 
imagine a motor controlled by a drive with speed demand of zero rpm. The motor shaft will 
appear immobile but if we apply torque to the shaft it will rotate through an angle that 
depends on the amount of torque applied. We can then apply the rated torque of the motor and 
then measure the “stiffness angle” in degrees. The resulting value provides an indication of 
the fidelity of the drive thus parametrised, although it is clearly not the only fidelity indicator. 
 
WHAT’S NEEDED 
 
To calibrate an HPD drive correctly we recommend using an oscilloscope with memory. 
Obviously, the technician in charge must be well versed in the use of the instrument. If it is 
impossible to use an oscilloscope, we illustrate, at the end of this section of the manual, an 
alternative although less accurate method.