4.6.2 Password Protected Menus
The customer settings sub-menus contain the parameters that may be customised by the end user. To prevent unauthorised changes being
made these menus are password protected and cannot be accessed until the password has been entered correctly.
4.6.3 Hour Meters
4.6.4 Fault Log
The following codes are used within the system:
Note. Any faults that are active when the power is switched off, and remain active when the power is re-applied, will cause a new entry to be
added into the fault log.
To enter the password from the main operating menu press and hold both the and keys for approximately 5
seconds until the menu changes to the password prompt as shown
The flashing cursor will be positioned over the first digit.
Using the key change the first digit of the code and press . The cursor will move to the next digit.
Repeat the process and enter the following password 1 2 1 _ _.
When the password has been entered correctly the Hour Meters menu will be displayed.
Use the key to navigate to page 3 “Customer Settings”.
The time in hours that the generator has been producing gas.
The time in operating hours that the generator can produce gas before a service is required.
The time in hours that the generator has been operating in Economy mode.
The fault log contains details of the ten most recent faults that have occurred on the generator. Each fault is
represented by a code which is displayed along with the time (hours run) at which the fault occurred and the
status of the fault.
Fault Codes Notes
C1 Pressure Start Inhibit Low inlet pressure. Inhibits start.
P1 Inlet Pressure Fault Low inlet pressure during cycling.
P2 Pressure Sensor Fault Pressure sensor communication error.
E1 Power Failure
Y1
O
2
Alarm
Y2
O
2
Communication failure Communication fault between O
2
analyser and control board
Y3 Incorrect cell selected
Y4
O
2
high (out of range) Occurs when O
2
> 25% (% generators) / O
2
> 1.05% (ppm generators)
Y5
O
2
zero drift error
Contact Parker domnick hunter
S1 Service due