18
5 Servicing
5.1 Cleaning
Clean the equipment with a damp cloth only and avoid excessive moisture around any electrical sockets. If required you may use a mild
detergent, however do not use abrasives or solvents as they may damage the warning labels on the equipment.
5.2 Service Intervals
The service operations should be performed at the hours run or fixed time intervals specified below (whichever occurs first).
- Check – Replace – Recommended Process
Description Of Service Required Typical Recommended Service Interval
Component Operation
Daily
Weekly
2000 Hours
(3-month)
4000 Hours
(6-month)
8000 Hours
(12-month)
16000 Hours
(24-month)
Generator Check status indicators located on the control panel.
Generator Check regulated outlet pressure
Generator
Check O
2
purity
System Check filter drains
O
2
Cell
Calibrate oxygen sensor
System Check inlet air quality
Generator Check for air leaks
Generator Check pressure gauges during purging for excessive back pressure
Generator Check condition of electrical supply cables and conduits
Generator Check cyclic operation
Generator
Replace mistx silencer
Recommended Service A
Filtration
Replace buffer tank filter.
Recommended Service B
Generator
Replace/Calibrate oxygen sensor
Recommended Service C
Generator
Replace/Service valves
Recommended Service D
A
B
C
A
B
with O
2
Analyser
without O
2
Analyser