C-7100 Series User Manual PEAK INSTRUMENTS INC.
13
5.Printer check
Check the printer interface device is working properly. The result is displayed correctly √, an error is displayed result
is ×, the buzzer alarm.
6.Tungsten lamp examination
Open the instrument tungsten light source, check the operating parameters of tungsten lights are working properly.
If the parameter is not working properly, then re-set the operating parameters of a tungsten lamp. Change
detections are always displayed correctly, the result is always √.
7.Deuterium lamp examination
Open deuterium light source instrument operating parameters to check the deuterium lamp is working correctly. If
the parameter is not working properly, then re-set the operating parameters of the deuterium lamp. Change
detections are always displayed correctly, the result is always √.
8.Signal detector check
Check signal detector instrument is working correctly. The result is displayed correctly √, an error is displayed result
is ×, the buzzer alarm.
9.Wavelength Calibration
Wavelength parameter checking instrument is working properly. Correct, then the pop-up boxes, please confirm
whether the user input wavelength calibration if no user input within five seconds, then skip this. If the argument is
wrong wavelength, then start looking for a deuterium lamp to automatically correct the characteristic peak
wavelength. Wavelength calibration result is displayed by √, the correction is not passed, then the result is ×, the
buzzer alarm.
10.Dark current correction
Read instrument dark current of energy, checking eligibility. If the dark current in the proper range, which means
that the dark current is correct, then displays the results of √. If the dark current exceeds the maximum setting, the
user is prompted dark current error. The results are displayed as ×, the buzzer alarm.
11.System parameters check
Instrument system baseline reading is correct. If correct, then the pop-up box that asks the user whether to re-enter the
correction system baseline, baseline correction system default does not automatically skip the 3 seconds. If an error does
not exist or the baseline, the baseline correction system directly. Showing results √, the correction is not
passed, then the result is ×, the buzzer alarm.
Self-test end, after re-calibration dark current, into the main instrument display.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Peak C-7100 Series and is the answer not in the manual?

Peak C-7100 Series Specifications

General IconGeneral
BrandPeak
ModelC-7100 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals