EasyManuals Logo

Peak C-7100 Series User Manual

Peak C-7100 Series
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
C-7100 Series User Manual PEAK INSTRUMENTS INC.
14
Note: After power on the instrument, the instrument will automatically self-test and initialization until after
initialization is complete, the instrument will warm up for 20 minutes, 20 minutes warm-up time or press [ESC] to
skip preheating case, the instrument being prompted preparatory work environment, which is the instrument re-
calibration dark current, set the working parameters, etc., and then enter the main menu.
V. Photometric measurement
1. Function description
Photometric measurements are measuring the absorbance of the sample at a single wavelength, transmittance or
energy value.
2. Set measuring mode
Press the set button, enter the measurement mode settings menu, select the desired test mode, press [ENTER] to
confirm. If you choose the energy model, energy window appears and prompts to select the amplifier gain.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Peak C-7100 Series and is the answer not in the manual?

Peak C-7100 Series Specifications

General IconGeneral
BrandPeak
ModelC-7100 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals