EasyManua.ls Logo

Peak C-7100 Series - Set Wavelength

Peak C-7100 Series
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-7100 Series User Manual PEAK INSTRUMENTS INC.
15
3.Set wavelength
Set the current working wavelength range 190nm-1100nm, press the up and down button to open wavelength
setting window, press [0] - [9] to enter the desired wavelength, press [CLEAR] to clear the input, press [ENTER] to
confirm. Input errors or exceeds the set range buzzer alarm.
4. Correction 100%T/0Abs
In two samples simultaneously into the two slots reference solution, and then press the [ZERO] key, the instrument will
be blank correction in the current wavelength. Display calibration is completed 100.0% T or 0.000Abs.
5. Measurement data
Calibrated with reference solution 100% T/0Abs, remove the rear slot reference sample solution into the sample solution,
and then press the [START] button, then perform a test, the sample data are immediately added to the list.

Table of Contents

Related product manuals