SGM700 Digitizer
First use of indicator -continue-
Calibration Settings -continue-
When a number is shown, the deletion of one calibration point is completed and
more points need to be deleted. Press key 3 >3 sec to do so.
>3 sec.
When all calibration points are deleted, the following screen is visible:
Entering new calibration points.
Use key 1 <2 sec. to go to CAL 1 and press key 2 <2 sec.
After entering, the following screen is visible:
And will automatically jump to:
First calibrate the zero point (CP1). Make sure the weigher is unloaded and
press key 3 >2 sec.