SGM700 Digitizer
Configuration Menu -PCL- -continue-
In PcL 1 you can set the polaritiy of the input range. Un=Unipolar mode the
input range for load cells is –0.2 mV/V to + value set at Range. Bi=Bipolar mode
the input range for load cells is –value set at Range to + value set at
Range.Press key 2 <2 sec. To change the polaritity og the input range.
The following screen is visible:
Use key 1 change the polarity. Confirm by pressing key 3 for >2 sec.
Up Down Confirm
The following screen is visible:
In PcL 2 you can set the Amplifier sensitivity. Choose between 1.0mV/V,
1.5mV/V, 2.0mV/V, 2.5mV/V and 3.0mV/V. Press key 2 <2 sec. to change the
amplifier sensitivity.
The following screen is visible:
55