Page 84 of 112 RPU 352, RPU 353, RPC 80, RPC 50
T6
This recording run can't be saved as a template. In order to save a recording run as a template
the following conditions must all be met:
• Channel 1 must have been active in the P.U. monitor.
• The run must contain no off-scale readings on any active channel.
• The run must contain at least 25 records.
This means that a run must be more than 4.2 minutes long with a recording interval 10
seconds or more than 12 minutes long with a recording interval of 30 seconds and pro-
rata for other intervals.
• There must have been no communications errors during the data transfer from the P.U.
monitor to the RPC-80.
• All the expected data must have been received from the P.U. monitor.
T7
Saving this template failed when the memory became full.
• Save the template in the place of a template that is no longer required.
• There may still be room in the memory for a smaller template.
Two sizes of template storage memory are available for the RPC-80, one with 127 sectors and
the other with 511. If you have the smaller type and full template memory becomes a problem
then you may wish to have your RPC-80 upgraded.
T8
Saving this template failed when a communications error occurred
• Connects the lead between the P.U. monitor and the RPC-80.
• Never unplug this lead until the "Template Saved OK" message shows.
• When a communications error occurs the partly saved template is deleted and another
an other chance will be offered to save the run as a template.
ATTENTION!
It is possible that an error could be caused by very severe interference
from other nearby equipment or on the mains wiring.