K23-SSI/USB/25B-C  | OPERATING SOFTWARE OS6.0 
45 
COUNTER B 
Mode “Counter”: Reset / Set the counter 
value of channel B to the set value in SET 
(s) 
A 
Mode „Counter“: Counter (channel A) is 
locked and does not count any further 
pulses as long as this command is present. 
B 
Mode „Counter“: Counter (channel B) is 
locked and does not count any further 
pulses as long as this command is present. 
Resetting the minimum / maximum value 
Device is reset to factory settings (pulse 
must be applied for at least one second!) 
INPUT 1 CONFIG 
This parameter determines the switching characteristics for "Ctrl. In 1." 
Activates at "LOW" (static) 
Activates at "HIGH" (static) 
Activates at rising edge (dynamic)  
Activates at falling edge (dynamic) 
INPUT 2 ACTION 
This parameter determines the control function of the input "Ctrl. In 2"  
See function assignment for parameter INPUT 1 ACTION 
INPUT 2 CONFIG 
This parameter determines the switching characteristics for "Ctrl. In 2." 
See activation assignment for parameter INPUT 1 CONFIG
 
INPUT 3 ACTION (FACTORY SETTINGS) 
This parameter is preset to "Factory Settings" (resets device to factory settings) 
and cannot be changed. 
INPUT 3 CONFIG (RISING EDGE) 
This parameter is preset to "Rising Edge" and cannot be changed.