28 . DSC 6000 Installation and Hardware Guide
Safety Precautions for the DSC 6000
WARNING
AVERTISSEMENT
Be sure that all instrument operators read and understand
the following precautions. It is advisable to post a copy of
these precautions on or near the instrument itself.
Assurez-vous que tous les opérateurs d'instruments lisent
et comprennent les précautions suivantes. Il est conseillé
de publier une copie de ces précautions sur ou près de
l'instrument lui-même.
The following precautions must be observed before and when using the DSC 6000
Differential Scanning Calorimeter:
• Before connecting the DSC 6000 to the main outlet, check the main voltage setting and
fuse.
• The DSC 6000 requires a good earth ground that is common to the earth ground of the
computer.
• Check that the power cord supplied is appropriate for your country and is undamaged
before connecting it to the main voltage supply.
• Use proper lifting posture when lifting the analyzer. The DSC 6000 with autosampler
weighs 19 kg. Lift the autosampler from underneath. Never attempt to lift the analyzer
by the autosampler (if installed) or by any cables attached.
• When cleaning the instrument, consult PerkinElmer if there is any doubt about the
compatibility of decontamination or cleaning agents with parts of the equipment or with
material contained in it.
WARNING
AVERTISSEMENT
Never touch the outer and inner furnace lids. The
temperature of the furnace can reach as high as 445 °C. Use
tweezers to remove the lids.
Ne touchez jamais les couvercles extérieurs et intérieurs du
four. La température du four peut atteindre 445 ° C. Utiliser
une pince à épiler pour enlever les couvercles.