AS 8000 Autosampler . 85
Error Messages from the DSC 6000
Error message ID Error Possible Cause Corrective
Action
ERR_DUE_TO_AUTOSAMPLER AS 8000 entered
fatal error state.
This caused DSC
to enter fatal
error state
Refer to AS 8000
error message
Refer to AS 8000
troubleshooting
(above)
ERR_AUTOSAMPLER_RESPONS Timeout on
command.
Response from
command
transmitted to AS
8000
I2C connection
cable defective
Replace I2C cable
ERR_AUTOSAMPLER_I2C_WD No link between
DSC and AS 8000
via I2C
I2C connection
cable defective
Replace I2C cable