DSC 6000 Hardware . 63
Subambient Operation of the DSC 6000
For cooling with liquids, many commercial circulating coolers are available that can be
interfaced to the DSC 6000. The Intracooler (P/N N5374098/N5374099) is recommended for
the range 65 °C to 450 °C. The Portable Cooling Device (P/N N5202068) can be used to
quickly cool the DSC 6000 to ambient temperature, or to cool below room temperature for
occasional subambient measurements. The PCD can be filled with ice, ice water, liquid
nitrogen or any other cooling mixtures.
The DSC 6000 with CryoFill system can be operated in the subambient mode in the range
170 °C to 300 °C. Liquid nitrogen serves as the coolant and nitrogen is used as the purge
gas to the sample area. It is also possible to use helium as a purge gas.
Subambient operation of the DSC 6000 is covered in the following topics in the Help:
Precautions for Subambient Operation
Operating the DSC 6000 in the Subambient Temperature Range
Connecting the DSC 6000 to the Dewar
Portable Cooling Device (PCD)