64 . DSC 6000 Installation and Hardware Guide
DSC 6000 Maintenance
The DSC 6000 needs little routine maintenance other than proper treatment as a sensitive
electromechanical device.
Since the sensor disk is the hottest component in the furnace area, it will usually remain
clean and free of condensates. However, samples may ooze out of sample pans or be
accidentally spilled onto the sensor disk. Such materials should be removed at low
temperatures whenever possible by using tweezers or moistened cotton swabs.
Solvents should not be placed directly on the sensor. Moisten a cotton swab or a tissue with
solvent; then clean the sensor or upper portion of the furnace wall with the swab.
Any residual sample that has carbonized may remain on the sensor disk after it has been
cleaned with a solvent. To remove the residue, gently scrub the sensor with a fiberglass or
nylon brush. Be careful not to exert excessive force when scrubbing the sensor since it can
be damaged. The sensor must be as clean and flat as possible for proper operation.
To avoid contaminating the furnace, use a purge gas at proper flow rates. If the furnace
becomes contaminated, clean the furnace as soon as possible.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer DSC 6000 and is the answer not in the manual?

PerkinElmer DSC 6000 Specifications

General IconGeneral
TypeDifferential Scanning Calorimeter
Temperature Resolution0.01 °C
Temperature Accuracy±0.1 °C
SoftwarePyris Software
Calorimetric Sensitivity0.2 μW
Power Requirements100-240 VAC, 50/60 Hz
Heating Rate0.1 °C/min to 100 °C/min
Cooling Rate0.1 °C/min to 100 °C/min
DetectorHeat Flux
AtmosphereInert, Oxidizing, Reducing
Sample Pan TypeAluminum, Platinum
Data Sampling RateUp to 100 points per second