30 . DSC 6000 Installation and Hardware Guide
CAUTION
ATTENTION
Care must be taken when handling the outer furnace lid so its ceramic
lining is not damaged.
Des précautions doivent être prises lors de la manipulation du couvercle
du four extérieur, de sorte que sa doublure en céramique n'est pas
endommagée..
Avoid contamination of the furnace, use high-purity purge gases to purge any reaction
products out of the furnace, and heat the furnace to temperatures above 430 °C ONLY
when absolutely necessary.
Check that the cooling liquid supply is circulating before starting a run.
Ensure the purge gases are set to the recommended flow rates. A rate that is too high
will disturb the inner furnace lid.
Do not operate the instrument in a cold room. The ambient temperature and the
temperature of the instrument should be between 10 °C and 30 °C.
Never open the furnace when it is at an elevated temperature.
Important Specifications
Maximum power consumption 430 VA
Supply voltage 100120 V or 220240 V
Supply voltage frequency 5060 Hz
Fuses 6.3 AT or 3.15 AT
Safe temperature range 1040 °C
Dimensions L x H x D 38 x 17 x 41 cm (H 29 cm with autosampler)
Weight 17 kg (19 kg with autosampler)