52 . DSC 6000 Installation and Hardware Guide
14. Enter the name of the reference material used (we recommend indium for this
calibration), the Expected
H value, the Measured H result just measured, and the
Weight of the reference material used.
15. Click the Save and Apply button. In the Save As dialog box, select the default.d6c
calibration file and click OK. Answer Yes to overwrite the existing file to save the new
calibration values.
16. Click Close to exit Calibration and send the new calibration values to the DSC 6000.
The DSC 6000 calibration is complete.
This completes the installation of the DSC 6000.