Routine maintenance
62
Note! Hold the tube near its end to avoid it kinking when you push it in!
Note! Don’t leave a gap between the ends of the two tubes inside the
silicone connector because this affects the output tube volume.
Note! The tube from Tip 1, which is to be connected to the tube from Pump
1, has a “sock” shrunk onto it to help you distinguish it from the tube from
Tip 2. Connections should be as in the table:
Tip configuration Pump1 Pump2
Real time Real time Not used
Pre Pre Not used
Real time and Pre Real time Pre
Pre and Post Pre Post