32 . Lambda 365 Users Guide
Preparing the Laboratory
The compact size of the Lambda 365 enables you to install the spectrophotometer in a
limited space.
Make Room for installation
The Lambda 365 System needs sufficient space for a computer, printer and selected
accessories such as the autosampler. The system dimensions are 491 mm (W) × 538 mm
(D) × 258 mm (H) [19.3 in. (W) 21.1 in. (D) 10.1 in. (H)].
WARNING
In view of the electronic components in the Lambda 365
System, it should be positioned well away from any
instruments emitting an electromagnetic field and any
environment that may cause vibration.
Compte tenu des composants électroniques dans le Système
Lambda 365, il doit être placé très loin de tous les autres
instruments émettant des champs électromagnétiques et
toutes les autres sources produisant des vibrations.
Power
A grounded AC power supply must be used. The Lambda 365 accepts 100 to 240 VAC at 50
Hz or 60 Hz.
WARNING
In order to protect the system from electrical noise and current
surges, please connect the power cable to a GROUNDED
OUTLET.
Afin de protéger le système contre les bruits électriques
parasites et les surtensions de courant, veuillez brancher le
câble d'alimentation à une prise d’alimentation mise à la terre.