EasyManuals Logo

PFlow Industries D116 User Manual

PFlow Industries D116
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Revision: 3.0.3
5.8 4~
20mA Analog Output Calibration
Note
Each flowmeter
has been calibrated strictly before leaving factory. It is unnecessary to carry
through this step except when the current value (detected while calibrating the current loop)
displayed in Window M58 is not identical with the actual output current value.
The hardware detect window must be activated prior to calibration
follows:
Press enter password
"115800"
next power on, this window will close automatica
Press to calibrate the current loop 4
mA
At the same time, press or
to adjust the display
at this point, the 4mA has been calibrated.
Then, press to calibrate the
current loop 20mA output
The results are automatically saved in
EEPROM and won
5.9 ESN
We provide the Flowmeter
with a unique electronic serial number to identify each Flowmeter for the convenience
of the manufacturer and customers. The ESN
Attention
Other Operation refers to "
6
Menu
^
0 Enter
^
^
^
Enter
D116
Series Ultrasonic Flowmeter
20mA Analog Output Calibration
has been calibrated strictly before leaving factory. It is unnecessary to carry
through this step except when the current value (detected while calibrating the current loop)
displayed in Window M58 is not identical with the actual output current value.
The hardware detect window must be activated prior to calibration
the Analog Output
. The procedure is as
"115800"
, then press
to activate the detect menu. With no effect to
next power on, this window will close automatica
lly as soon as the power is turned off.
mA
output. Use an ammeter to measure the current loop output current.
to adjust the display
ed numbers.
Watch the ammeter until it reads
current loop 20mA output
. The method is the same as 4mA
calibration.
EEPROM and won
t lose when power off.
with a unique electronic serial number to identify each Flowmeter for the convenience
of the manufacturer and customers. The ESN
is able to be viewed in Window M61.
6
.2 Windows Display Explanations".
Enter
Series Ultrasonic Flowmeter
Page 21 of 53
has been calibrated strictly before leaving factory. It is unnecessary to carry
through this step except when the current value (detected while calibrating the current loop)
. The procedure is as
to activate the detect menu. With no effect to
output. Use an ammeter to measure the current loop output current.
Watch the ammeter until it reads
4.00. Stop
calibration.
with a unique electronic serial number to identify each Flowmeter for the convenience

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PFlow Industries D116 and is the answer not in the manual?

PFlow Industries D116 Specifications

General IconGeneral
BrandPFlow Industries
ModelD116
CategoryMeasuring Instruments
LanguageEnglish