Revision: 3.0.3
5 Operating Instructions
5.1 System No
rmal Identification
Press the keys. If the letter "
*R
If the letter "G"
is displayed, it indicates that system is adjusting the signal gain prior to the measurement. Also, it
means system normal. Only when the adjustment takes too long without stopping, can
abnormal.
Letter "I"
indicates no signal is being detected. Check the transducer wiring connections are correct, the
transducers are installed firmly, etc.
5.2 Zero Set Calibration
Once zero flow occurs, a zero point may indicate on ea
value is not equal to "0", this value indicates "Zero". To any measuring instrument, the smaller the "Zero"
better the quality is. Conversely, if the Zero is too big, that indicates the quality o
If the zero set point is not at true zero flow, a measurement difference may occur. The smaller the physical
measurement capacity is, the larger the measurement difference from the zero point will exist. Only when zero
point reduc
ed to a definite degree, as compared with the physical measurement capacity, can the measuring
difference from zero point be ignored.
For an ultrasonic Flowmeter, the measurement difference from zero point cannot be ignored at low flow. It is
necessary to perform a static
zero set calibration to improve low flow measurement accuracy.
Press Window M42 to set the Zero, press
corner reducing to be "0". If this is carried out with flow, the flow will be displayed as "0", M43 can help to
restore settings.
5.3 Scale Factor
Scale factor refers to the ratio between "
actual
and
is 2.00, and it is indicated as 1.98 on the instrument, the scale factor reading is 2/1.98. This means that the
best scale factor constant is 1.
However, it is difficult to keep the scale fa
ctor as
difference is called "consistency".
During operation, there still exists possible difference in pipe parameters, etc. The
when used on different pipes. Thus
, scale factor calibration is specially designed for calibrating the differences
that result from application on different pipes. The scale factor entered must be one that results from actual
calibration.
The scale factor can be input in Window M45.
5.4 System Lock
System lock is readable but not modifiable
to prevent operation error due to unauthorized tampering by
unauthorized personnel.
Press the keys,
if displays "Unlock" on the screen, then
long password, and then press the
key to confirm.
Unlock it by
using the selected password only. Press
press the key and
enter the correct password, then press
Keep the password in mind or reco
rded in a safe place
Menu 0 8
Enter
Menu 4 7
Enter
Enter
D116
Series Ultrasonic Flowmeter
rmal Identification
*R
"
displays on the screen, it indicates system normal.
is displayed, it indicates that system is adjusting the signal gain prior to the measurement. Also, it
means system normal. Only when the adjustment takes too long without stopping, can
system be identified as
indicates no signal is being detected. Check the transducer wiring connections are correct, the
Once zero flow occurs, a zero point may indicate on ea
ch measuring instrument,
but the displayed measuring
value is not equal to "0", this value indicates "Zero". To any measuring instrument, the smaller the "Zero"
better the quality is. Conversely, if the Zero is too big, that indicates the quality o
f the instrument is poor.
If the zero set point is not at true zero flow, a measurement difference may occur. The smaller the physical
measurement capacity is, the larger the measurement difference from the zero point will exist. Only when zero
ed to a definite degree, as compared with the physical measurement capacity, can the measuring
For an ultrasonic Flowmeter, the measurement difference from zero point cannot be ignored at low flow. It is
zero set calibration to improve low flow measurement accuracy.
first, and then wait the readings displayed at the lower right
corner reducing to be "0". If this is carried out with flow, the flow will be displayed as "0", M43 can help to
actual
value" and "reading value"
. For example, when the measur
is 2.00, and it is indicated as 1.98 on the instrument, the scale factor reading is 2/1.98. This means that the
ctor as
"1" on the instrument especially in batch
productions
During operation, there still exists possible difference in pipe parameters, etc. The
"scale factor"
may be necessary
, scale factor calibration is specially designed for calibrating the differences
that result from application on different pipes. The scale factor entered must be one that results from actual
The scale factor can be input in Window M45.
to prevent operation error due to unauthorized tampering by
if displays "Unlock" on the screen, then
press the
key, enter
key to confirm.
using the selected password only. Press
, if "lock" is displayed on the
screen
enter the correct password, then press
to confirm.
rded in a safe place
, otherwise the instrument cannot be used.
Enter
Enter
Menu 4 7
Enter
Series Ultrasonic Flowmeter
Page 19 of 53
displays on the screen, it indicates system normal.
is displayed, it indicates that system is adjusting the signal gain prior to the measurement. Also, it
system be identified as
indicates no signal is being detected. Check the transducer wiring connections are correct, the
but the displayed measuring
value is not equal to "0", this value indicates "Zero". To any measuring instrument, the smaller the "Zero"
is, the
f the instrument is poor.
If the zero set point is not at true zero flow, a measurement difference may occur. The smaller the physical
measurement capacity is, the larger the measurement difference from the zero point will exist. Only when zero
ed to a definite degree, as compared with the physical measurement capacity, can the measuring
For an ultrasonic Flowmeter, the measurement difference from zero point cannot be ignored at low flow. It is
first, and then wait the readings displayed at the lower right
corner reducing to be "0". If this is carried out with flow, the flow will be displayed as "0", M43 can help to
. For example, when the measur
ement
is 2.00, and it is indicated as 1.98 on the instrument, the scale factor reading is 2/1.98. This means that the
productions
. The
may be necessary
, scale factor calibration is specially designed for calibrating the differences
that result from application on different pipes. The scale factor entered must be one that results from actual
to prevent operation error due to unauthorized tampering by
key, enter
6 numerically
screen
, then

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PFlow Industries D116 and is the answer not in the manual?

PFlow Industries D116 Specifications

General IconGeneral
BrandPFlow Industries
ModelD116
CategoryMeasuring Instruments
LanguageEnglish