Revision: 3.0.3
10.4
Transducer Mounting Methods
W211
insertion transducer mounting method: Z method through M24, it should be installed according to the
specific application condition.
10.4.1 Z Mounting Method
Z method is the most commonly used mounting met
diameters ranging from 50mm to 12
00mm. Due to strong signal strength and high measurement accuracy, the Z
method is preferable for pipe sections severely rusted or with too much scale formatio
installing the transducer using the Z method, be sure that the two transducers and the pipeline center axis are in
the same plane, but never in the 6 or 12 o’
clock positions. see below:
10.4.2
Pipe Parameter Entry Shortcuts
For examp
le, measuring the diameter of DN200, pipe outside diameter is 219mm, pipe wall thickness is 6mm,
pipe inner diameter is 207mm, measuring medium is water, and material is carbon steel, no liner, can be operated
as follows:
Step1. Pipe outside diameter:
Press
M11 and enter the pipe outside diameter, and
then press the key to confirm.
(
For insertion transducer, M11 menu need to be
entered the pipe inner diameter)
Step2. Pipe wall thickness
Press the
window M12, and enter the pipe wall thickness,
and press the key to confirm.
(The Wall Thickness needs to be 0.01mm for
use with insertion sensors.)
Step3. Pipe Material
Press the
window M14, press the key, p
or
key to select Pipe Material, and
press the key to confirm.
Profile Chart
Menu
1 1
Enter
Menu
1
2
Enter
Menu
1
4
Enter
^
^
Enter
D116
Series Ultrasonic Flowmeter
Transducer Mounting Methods
insertion transducer mounting method: Z method through M24, it should be installed according to the
Z method is the most commonly used mounting met
hod for insertion-
type ultrasonic flowmeters, suitable for pipe
00mm. Due to strong signal strength and high measurement accuracy, the Z
method is preferable for pipe sections severely rusted or with too much scale formatio
n on the inside wall. When
installing the transducer using the Z method, be sure that the two transducers and the pipeline center axis are in
clock positions. see below:
Pipe Parameter Entry Shortcuts
le, measuring the diameter of DN200, pipe outside diameter is 219mm, pipe wall thickness is 6mm,
pipe inner diameter is 207mm, measuring medium is water, and material is carbon steel, no liner, can be operated
M11 and enter the pipe outside diameter, and
For insertion transducer, M11 menu need to be
window M12, and enter the pipe wall thickness,
(The Wall Thickness needs to be 0.01mm for
key to select Pipe Material, and
Reference point
At 9 o’clock
Top View
Drill hole spacing
L
Flow
207 mm
0.01 mm
0. Carbon Steel
Series Ultrasonic Flowmeter
Page 49 of 53
insertion transducer mounting method: Z method through M24, it should be installed according to the
type ultrasonic flowmeters, suitable for pipe
00mm. Due to strong signal strength and high measurement accuracy, the Z
n on the inside wall. When
installing the transducer using the Z method, be sure that the two transducers and the pipeline center axis are in
le, measuring the diameter of DN200, pipe outside diameter is 219mm, pipe wall thickness is 6mm,
pipe inner diameter is 207mm, measuring medium is water, and material is carbon steel, no liner, can be operated
Reference point
At 3 o’clock