EasyManua.ls Logo

Pixsys ATR142 - Page 16

Pixsys ATR142
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
To use the LATCH ON function configure as you wish the parameter
.
3
For the calibration procedure refer to the following table:
Press Effect Operation
1
F
N
C
Exit parameters
configuration. Display 2
shows the writing .
Position the sensor on the
minimum functioning value
(associated with )
2
Set the value to minimum.
The display shows
Position the sensor on the
maximum functioning
position (associated with
)
3
Set the value to maximum.
The display shows
To exit the standard
procedure press
F
N
C
.
For “virtual zero” settings
position the sensor on the
zero point.
4
S
E
T
Set the virtual zero value.
The display shows
N.B.: for selection of
the procedure in
point 4 should be followed
on each re-activation.
To exit the procedure press
F
N
C
.
MAX
MIN
ZERO
3
The tuning procedure starts by exiting the configuration after changing the parameter.

Related product manuals