5 FOR YOUR SAFETY
User’s manual Planmeca Chair 5
5 FOR YOUR SAFETY
When driving the chair from one position to another care
should be taken. In case of emergency the chair will stop
when the backrest, the stop plate underneath the seat, or
the stop bar or stop cover on the underside of the lifting
mechanism is pressed upwards. The chair will also stop
when the safety switch on the underside of the legrest is
activated. The chair can be driven normally after the
possible obstruction has been removed.
Because of the chair swivel feature the stop plate located
on the underside of the seat does not cover the whole
surface.
Electromagnetic interference between the equipment
and other devices can occur in very extreme condi-
tions. Do not use the equipment in close conjunction
with sensitive devices, or devices creating high elec-
tromagnetic disturbances.
Switch off the unit before using an electrosurgical
knife.
Do not use the equipment in close conjunction with
anaesthetic gas or in highly oxygenated environ-
ments (oxygen content >25%).
No modification of this patient chair is allowed.
To avoid risk of electric shock, Planmeca Chair must
only be connected to a supply mains with protective
earth.
Canada only:
External fuse: max 20A branch circuit fuse.
Do not perform maintenance procedures or replace
patient chair parts while treating a patient.
PM_Chair1.eps
Stop bar
Stop plate
Backrest
Safety
switch
Stop cover