EasyManua.ls Logo

Primes PowerLossMonitor PLM Series - Analog output signal description; Installation of the external temperature sensor; Temperature sensor installation guide

Primes PowerLossMonitor PLM Series
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
Revision 01/2019 EN
PowerLossMonitor PLM
35,0(6
8.10 Analog output
The analog signal is effected via the 4-pin device socket M8 (see Fig. 8.2 on page17 or Fig. 8.4 on
page18 and Fig. 8.5 on page19). The output voltage is 0V ...10V. A suitable connection cable is
included in the scope of delivery.
PLM 2 PLM 10 PLM 20 PLM 40
An output voltage of 1V equals approx. 250W 1000W 2000W 4000W
Tab. 8.3: Output voltage and laser power
Pin assignment socket
(top view, plug-in side)
Pin Wire color Function
1 Brown Not connected
2 White Not connected
3 Blue Ground for the analog signal
4 Black Analog signal 0-10 Volt
Tab. 8.4: Socket assignment of the analog output
9 Installation of the external temperature sensor
The temperature sensor has to be installed in a way that ensures that it extends into the water flow but
against the flow direction. The tip should protrude from the bottom of the pipe approximately 15mm.
1. Plug the external temperature sensor into the cooling circuit output using a T-piece (see Fig. 7.4 on
page16).
X The temperature sensor should be positioned as close to the cooling circuit output of the mirror or the
absorber as possible.
Flow Direction
Fig. 9.1: Installation of the external temperature sensor

Table of Contents

Related product manuals