7.3 Humidity
The device must not be operated in a condensing atmosphere. The humidity has to be considered in
order to prevent condensates within and outside the device.
The temperature of the cooling water must not be lower than the dew point (see Tab. 7.1 on page14).
NOTICE
Damage/Destruction of the device due to condensing water
Condensation water inside of the device can lead to damage.
X
Mind the dew-point in Tab. 7.1 on page14.
Do only cool the device during the measuring operation. We recommend starting the cooling approx. 2min-
utes before the measurement and terminating it approx. 1 minute after the measurement.
Cooling water temperature in °C
0
5
10
15
20
25
30
35
40
0 5 10 15 20 25 30 35 40
Air temperature in °C
100
50
60
70
80
40
30
20
10
Relative humidity in %
Tab. 7.1: Dew point diagram
Example
Air Temperature: 22 °C
Relative Humidity: 60 %
The cooling water temperature cannot fall below 14 °C.