MR-20 OPERATOR’S GUIDE
47
2. Make sure the Pathway
®
Electrode (Part #6750) is snapped properly into the
Pathway
®
Preamplier (Part #2583) and is rmly attached to the forearm.
If the cable passes the rst test, and you still have high or unstable readings, try a
new electrode. This may solve the problem.
For further assistance, please contact The Prometheus Group
®
at 1-800-272-8495.
3. Test the Pathway
®
Adapter (Part #3660):
If the Pathway
®
Preamplier (Part #2583) passes the previous tests,
and readings are still erratic, try replacing the Pathway
®
Adapter (Part
#3660). Reconnect device, accessories and consumables, and if the
readings continue to be unstable, it could be the adapter.
G. Repeat steps A through H using the second Pathway
®
Preamplier
(Part #2583) and EMG B.
H. If either Pathway
®
Preamplier (Part #2583) shows signs of instability
while you are performing the “stress test” then the preamplier cable
fails and would require replacement.
Erratic Readings (Continued)
4. Test the Sensor (Pathway
®
Intracavity Sensors Part #6330, Part #6340, and
Part #6320)
If the Pathway
®
Preamplier (Part #2583) passes the previous tests, and
readings are still erratic, try replacing the Pathway
®
Intracavity Sensor(s)
(Part #6330, Part #6340, or Part #6320). Reconnect device, accessories and
consumables, and if the readings continue to be unstable, then the sensor was
causing the issue.