COMPACT 2 NDT
User Manual
519
0
-
0
-
0002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
15
von
65
1. Mains 220-240V, 16 A
Cu wire-cable 2.1,5mm², Ground-Leakage-Switch 25A / 30mA; Connected load 2.2 kW..
2. 2a Power switch:16 A, 140cm above floor.
2b Machine connection: Socket incl. Earth 60cm above floor.
3. 3a Water connection: Water consumption 1,9l/min. Water temperature may not be below 5°C.
3b Waters top cock ¾“connection.
4. Ventilation of darkroom is necessary.
5. Drainage hose PVC, diameter 50mm, acid resistant. Odour lock with hose connection.
6. Drainage for developer, Hose connection 10mm. Suitable collection containers must be
provided. Pay attention to local wastewater regulations!
7. Drainage for fixer; Hose connection 10mm. Suitable collection containers must be provided.
Pay attention to local wastewater regulations!
8. Sink with hot/cold water and flexible hose; Inner dimensions are 60x40x30cm deep. Material:
Ceramic, stainless steel, plastic.
9. The replenishment containers can be place under the processor or can be installed externally.