Protec COMPACT 2 NDT
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMPACT 2 NDT
User Manual
519
0
-
0
-
0002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
28
von
65
3 Now by pressing one of the buttons 4 or 5 change the display: “L.0” if you want to unlock the
input or “L.1” if you want to lock the input.
4
To finalize the changes press button 6 “Exit”.
NOTE
Even when the input lock is activated it is possible to recall (previously stored)
processing parameters of the memory-function.
CAUTION!
Please note, if input is locked via lock function, then processing parameters
cannot be stored - when trying to do this, an error message “LO” will be
displayed.
5.2.16
Stop Film Transport
In a case of a film-jam inside the machine, the film transportation can be manually interrupted. To stop
the film transport press both arrow-buttons (2 and 3) simultaneously.
Related topics:
See “Manual starting and stopping the film transport: 5.2.13.3
See “Film is caught up in the racks”: see 5.4.2.10
5.2.17 Infrared Sensitive Materials
Infra-red sensitive film material can be processed without danger of exposure by the light-barrier.
5.2.18 Automatic Cooling Function
If the developer temperature exceeds the set value, the automatic cooling function will be activated.
The fan runs for a fast cooling of the bath. The same happens when the temperature is decreased by
pressing the arrow button.