5.4.2 Advice on Machine Errors
5.4.2.1 Machine does not switch on
Ensure that electrical plug is firmly inserted into socket. Ensure that electrical socket has power
supply by testing with an appliance (e.g. table light)
5.4.2.2 Rinsing water does not flow
Open water inflow tap
Water pressure in the water system is too low: Minimum pressure 2 bar (29 psi)
5.4.2.3 Replenishment rate too high
Check the programmed times of replenishment cycle and replenishment time
5.4.2.4 Replenishment rate too low
Check the programmed times of replenishment cycle and replenishment time
Check whether the machine can be started by each of the light barriers (4 pieces). In the case
of a light sensor being faulty, call your service technician. In the meantime, double the
programmed replenishment time
5.4.2.5 Replenishment pump does not pump
Check whether the replenishment containers are full and that the end of the suction pipe is
positioned under the liquid level
Check whether there is air in the replenishment pipes. If this is the case, then check the pipe
connection
5.4.2.6 Water tank overflows
Water drainage hose (overflow) is bent. The hose end should be positioned above the drainage
level in the syphon
Check water drainage in the tank and hose for blockage and pollution. The drainage hoses
should have a constant inclination
5.4.2.7 The film does not transport correctly
Film is fed in and gets caught in the machine: Check the positioning of the racks in the
machine and make sure that the latches are closed.
5.4.2.8 Film cannot be fed, the display shows „E1“
Close machine cover securely, paying special attention that the switch on the rear of the
control panel is actuated
5.4.2.9 Film is in the infeed and nothing happens
Display is flashing: Control is in the manual mode. Switch back to Automatic mode
(see 5.2.2 )
Developer-temperature button is flashing: The temperature of the developer has not yet been
reached. Wait until temperature has been reached and the button stops flashing
Push the film deep into the infeed and move this from side to side. If the processor still does
not start then call your service technician. In the meantime, with limitations the processor can
be worked with in the Manual Mode
5.4.2.10 Film is caught up in the racks
Switch machine off and remove cover
Check in which rack the film is caught up in and remove the respective rack
If possible, catch hold of the film end by hand and by manually turning the drive-shaft, pull the
film out of the rack
Replace the rack and secure with fastener. Replace machine cover and switch the machine on
again.
If a film is caught up in the machine due to a power cut, it can be transported out of the
machine by activating the transport with a film in the infeed (machine can also be started up in