Page 28 
P605 Volume Units 
This parameter determines the units that will be used in calculating volume 
conversion. It is used in conjunction with P607 (maximum volume), please 
note that there is no provision for the volume unit’s descriptor to be shown 
on the display.  
The choices are: 
Volume will be totalised with no units 
Volume will be totalised in Tons 
Volume will be totalised in Tonnes 
Volume will be totalised in cubic metres 
Volume will be totalised in litres 
Volume will be totalised in UK Gallons 
Volume will be totalised in US Gallons 
Volume will be totalised in cubic feet 
Volume will be totalised in barrels 
 
P606 Correction Factor 
This parameter is used to enter a correction factor, when required, such as 
the specific gravity of the material so that the volume calculated is relative 
to the actual amount of material that can be contained between empty level 
(P105) and 100% of span (P106). Default = 1 
P607 Max Volume 
This  parameter  displays  the  actual  maximum  volume  that  has  been 
calculated by the IMP+ i.s., i.e. P604 Calculated Volume x P606 Correction 
Factor, and is a Read Only parameter. The volume displayed will be shown 
in P605  Volume Units and is  the total volume available  between  empty 
level (P105) and 100% of span (P106).