26. February 2024 © Copyright 2024, PROCEQ SA 48
5.16 P-wave estimation from measured S-wave value
Many users will be more familiar with P-wave values as an indicator of concrete
quality.
The P-wave velocity can be calculated directly from the S-wave velocity if the
Poisson’s ratio of the concrete is known.
Where :
V
P
= P-wave velocity
V
S
= S-wave velocity
ν = Poisson’s ratio
If the Poisson’s ratio is known this can be entered manually.
To simplify things, typical values for the Poisson’s ratio can be used based on
the compressive strength.
Figure 49: S-wave to P-wave velocity.
=
2
2
(1)
21