6. Troubleshooting
The PureWave system may recognize error situations on its own and show the corresponding messages on the
display. Some of these errors may be corrected by the user, without outside help. The following description
classifies errors roughly as those which may be shown on display with a comprehensive message and others
which may not be described, if only incomprehensibly.
For all errors the symptoms – and if possible – also the cause and the remedy are described.
In case of any doubt please contact the manufacturer, an authorized service partner or your
distributor! For all actions reaching beyond the measures as described in chapters 2 and 3
the AC cord has to be removed from the socket or the device’s jack!
6.1 Error messages on the display
No applicator has been connected or applicator connector
is loose.
Ü Please connect an applicator.
The applicator does not conduct current. The applicator or
its connection cable may be damaged, usually caused by
rough or improper handling.
Ü Please check if the applicator has been connected and tightened correctly. If available test another
applicator on the device’s corresponding connection. In case of visible (mechanical) damages, please contact
your service partner or distributor.
The control unit has detected an error during its self-test.
Ü Turn the control unit off and back on after a few seconds. If this is a permanent or recurring error, please
contact your service partner. Please note the error code >>nn<< and forward it to your service partner!