PureWave PEMI User Manual | 06-23 V1.0 Page 23 from 28
PureWave PEMI User Manual
6.2 Error codes during timer therapy
If a timer therapy (activated by the user via timer programming) could not be performed a fault report will
appear on the display the next time the device is started. This report must be acknowledged by pressing the
Start/Stop-button TO RESET. For each possible time-controlled therapy – timer 1 and timer 2– the display will
show the following messages:
In the last stated message the >>N<<states the number of the timer or therapy, >>xyz<< represents an error
code, meaning the following:
E101 No applicator has been connected.
Ü Please connect an applicator respectively the one prescribed on the chip card.
E102 The applicator does not conduct current. The applicator or its connection cable may be defect, usually
caused by
rough or improper handling.
Ü Please check if the applicator has been connected and tightened correctly. If available test another
applicator on the device’s corresponding connection. In case of visible (mechanical) defects please
contact your service partner or distributor.
E103 The therapy could not be started since the device was already in operation.
Ü Please check the starting times, correct them, if required and repeat the therapy, if necessary.
E104 The therapy could not be started due to an error within the device (communication problem).
Ü Please contact your service partner.
6.3 Further error situations
The device cannot be turned on, the display stays
dark and empty
Please check if the socket and AC cord conducts line
voltage, if this is the case check the AC fuse (see
chapter 2.2.1). If the problem persists, please
contact your service partner.
A permanent alarm signal resounds after connecting
the device to the AC power supply.
The control unit’s program memory has the wrong
content. Please contact your service partner.
The menu to set the time appears every time the
device is turned on.
The control unit’s clock is powered by a battery (its
regular life-span averages more than 5 years). The
battery is empty and must be replaced. Please
contact your service partner.