PureWave PEMI User Manual | 06-23 V1.0 Page 24 from 28
PureWave PEMI User Manual
7. Maintenance
Functionality, reliability, and safety characteristics of the device are only guaranteed in case of proper use in
accordance with the operating instructions. Safety control, maintenance work, repair work, and modifications
shall be carried out only by the manufacturer or by authorized service agents authorized. In case of a failure,
parts which influence the safety of the device may only be replaced by original spare parts of the manufacturer.
The electric installation must correspond to the requirements in accordance with VDE/IEC. The device does not
contain any parts which need maintenance work done by the user.
7.1 Safety Controls
7.1.1 Statutory requirements and regulations
Performance of technical safety checks is not obligatory, but is recommended by the manufacturer.
Pay attention to any national regulations at the place of installation/operation.
7.1.2 Performance of safety checks
It is recommended to have the device checked by the manufacturer or by a authorized service agency at 24-
months intervals.
The check shall consist of at least the following criteria:
• Electrical safety check in accordance with the test plan of the manufacturer
• Check of the device in respect of external integrity
• Check of all display and operating elements in respect of damages
• Check of all inscriptions in respect of legibility
7.2 Cleaning, Disinfection and care
For cleaning and disinfection of the device and its accessories there should not be used any agents containing:
higher portions of phenol derivatives, alcohol, compounds of chlorine or peracetic acid. It is recommended to
use disinfectants on aldehyde basis.
We recommend cleaning the device with a soft, damp cloth. For tenacious stains you may also apply mild
detergent used for delicate plastic surfaces.
The device is not suited for heat sterilization or for sterilization with gases.
Before cleaning or disinfection unplug the power supply out of the socket!
The device is suited for wiping disinfection. It has to be observed that no liquids enter the device. Under no
circumstance the plug or socket should get wet. For cleaning or disinfection the device may not be sprayed.
Control Unit, Applicators
Do not immerse in water! Use normal household cleaners for synthetic materials. Only wipe with a damp cloth.
Be sure to keep sharp, pointed objects away from mat applicator and pillow applicator. When using the coil mat
and coil pillow, please make sure that no sharp objects are pressed in. Do not expose the system to direct
sunlight and protect it from frost.
Placing Mat under a Mattress
The mat can be placed under a Mattress Pad, Mattress Protector, or woolen blanket that is placed against the
mattress and or under the sheets of the bed….in order to avoid undesirable bacteria or fungi caused by sweat.