PureWave PEMI User Manual | 06-23 V1.0 Page 4 from 28
PureWave PEMI User Manual
Notes
Guidance and manufacturer’s declaration electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in
such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Portable and mobile RF communications equipment should be used no
closer to any part of the equipment, including cables, than the
recommended
separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
Conducted RF,
IEC 61000-4-6
3 Veff
150 kHz to 80 MHz
3 Veff
d=1.2ÖP
Radiated RF,
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
d=1.2ÖP
for 80 MHz to 800 MHz
d=2.3ÖP
for 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts according to the transmitter manufacturer and d is the
recommended
separation distance in meters (m). Interference may occur in the
vicinity of equipment marked with the following symbol: