Use of this equipment adjacent to or stacked with other equipment should be avoided.
Ultraviolet Radiation!
To reduce the risk of UV Exposure:
Do not attempt to open or disassemble Sofia 2.
Do not attempt to look inside Sofia 2 while operating.
Failure to follow these warnings will invalidate the warranty.
Potential Biohazard!
To reduce the risk of biohazard:
Dispose of used specimens in accordance with Federal, State, and Local requirements.
Treat specimens and patient samples as potentially biohazardous material.
Ensure Sofia 2 is cleaned per the Maintenance and Cleaning section.
Seek specific training or guidance if you are not experienced with specimen collection and
handling procedures.
Use of Nitrile, Latex, or other gloves is recommended when handling patient samples.
Caution!
To reduce the risk of incorrect results:
Sofia 2 should only be used by trained operators.
Do not use if Sofia 2 is reporting an error condition that cannot be corrected.
To obtain accurate results, refer to the assay-specific Package Insert for details on specific tests.
Use the test kit within the expiration date.
To reduce the risk of Sofia 2 damage:
Sofia 2 is designed for countertop operation.
Sofia 2 is not designed to be stacked on top of each other.
Do not obstruct the flow of air on each side of the unit.
Maintain 6” minimum clearance on all sides including the back of the device.
Sofia 2 is not designed to withstand moisture, extreme humidity, or extreme temperatures.
Sofia 2 is not designed to withstand severe shock or vibration.
Do not open or disassemble the device.
Failure to follow the precautions mentioned above will invalidate the warranty.