Brief Warnings, Precautions, and Limitations
Always operate Sofia 2 on a surface that is level, dry, and not in direct sunlight.
Do not stack or place objects on top of the Sofia 2. Maintain a 6” minimum clearance
around the sides and back of the Sofia 2.
The Sofia 2 Calibration Cassette is sensitive to light. Store it in the opaque sleeve whenever
it is not being used.
Do not write on the barcode or surface of the Cassette.
Never move Sofia 2 or open the drawer while there is a test in progress.
Use only the Power Adapter that was provided with Sofia 2.
Do not drop Sofia 2 as it could damage the unit.
Contact Technical Support if the instrument is damaged.
Sofia 2 has been designed to CISPR 11 Class A. In a domestic environment, it may cause
radio interference, in which case you may need to take measures to mitigate the
interference.
Safety Precautions
Sofia 2 is designed to provide safe and reliable operation when used according to this User
Manual. If Sofia 2 is used in a manner not specified in the User Manual, the protection provided
by the equipment may be impaired.
All warnings and precautions should be followed in order to avoid unsafe actions that could
potentially result in personal injury or damage to the device.
Warning!
To reduce the risk of electrical shock:
Unplug Sofia 2 before cleaning.
Plug the device into an approved receptacle.
Do not immerse in water or cleaning solutions.
Do not attempt to open the enclosure.
Note: Significant Electrostatic Discharge (ESD) event(s) may permanently disable ethernet
port functionality. If this occurs, please use Wi-Fi or LTE (if available) as alternative
methods of communication or contact Customer Service.
To reduce the risk of improper operation:
Use of any cable or adapter other than those provided by the manufacturer could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment.
Ensure the device has a minimum separation distance of 12” for Portable RF
communications equipment (including peripherals such as antenna cables and external
antennas).

Table of Contents

Other manuals for Quidel Sofia 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quidel Sofia 2 and is the answer not in the manual?

Quidel Sofia 2 Specifications

General IconGeneral
Product NameQuidel Sofia 2
CategoryMeasuring Instruments
TypeFluorescent Immunoassay Analyzer
TechnologyFluorescent Immunoassay
ConnectivityUSB, Ethernet
DisplayTouchscreen LCD
Operating Temperature15°C to 30°C
Sample TypesNasopharyngeal swabs, nasal swabs
ResultsQualitative
Read Time15 minutes
Power Requirements100-240V AC, 50/60 Hz
Humidity Range10% to 90% non-condensing