EasyManuals Logo

rada DRV40 User Manual

rada DRV40
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
32
IMPORTANT! The DRV40 is locked in disinfection mode until one of the
following actions has been performed:
a. The 'Disinfection Timeout' period has expired (automatic).
b. The 'Abort'* button is pressed (manual).
Disinfection Timeline
Trigger
(Disinfection Timeout start)
Full Hot
Disinfection Timeout Period
Draw-Off Blend Circuit
Cool Down
return to Setpoint
(Disinfection Timeout nish)
*Abort
If Abort is used when the cycle is at full hot, the DRV40 switches to full cold and
displays "Error Temp". Cycle the power off/on to reset the DRV40, make sure the
blend circuit is at a safe temperature before allowing any outlets to be used.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the rada DRV40 and is the answer not in the manual?

rada DRV40 Specifications

General IconGeneral
Brandrada
ModelDRV40
CategoryControl Unit
LanguageEnglish