48 Chapter 3
Create an electrode
1. Press 4 Electrode Library, then 1 New electrode.
2. Press in the Function eld and select the function of the
electrode, refer to the table below:
Electrode type Select function
pH single pH
pH combined w/wo temp. sensor pH
Single metal/redox mV (i=0)
Combined metal/redox w/wo temp. sensor mV (i=0)
ISE single mV (i=0)
ISE Combined w/wo temp. sensor mV (i=0)
Reference single Reference
Temperature sensor T°C
Ground metal Ground
Double metal mV (i > 0)
3. Press in the ID eld.
4. Select Other in the From eld.
The option From = Catalogue allows you to create an electrode
from a list of Radiometer Analytical electrodes.
5. Enter the identication (ID) of the electrode, (up to 16 characters).
6. Press 1 twice.