24 Chapter 2
Apply internal stirringApply internal stirring
Apply internal stirringApply internal stirring
Apply internal stirring
1. For Instrument, select the stirrer of the Titration Manager or the
stirrer of an ABU62 connected to the Titration Manager.
2. Select Internal stirring = ON,
3. Press ✓ in the field Speed and select a stirring speed.
Apply external stirringApply external stirring
Apply external stirringApply external stirring
Apply external stirring
1. Connect the Stirring Propeller, part no. 847-731, to the Titration
Manager or ABU62 Propeller socket. Line 3 is automatically
replaced by External stirring.
2. For Instrument, select the stirrer of the Titration Manager or the
stirrer of an ABU62 connected to the Titration Manager.
3. Select External stirring = ON.
4. Adjust stirring by turning the stirring propeller knob. You can consult
the corresponding table between the position (1 to 9) and the stirring
speed by pressing ✓ in the field Speed setting.
Fill the Karl-Fischer cell (TIM980 only)Fill the Karl-Fischer cell (TIM980 only)
Fill the Karl-Fischer cell (TIM980 only)Fill the Karl-Fischer cell (TIM980 only)
Fill the Karl-Fischer cell (TIM980 only)
1. For Instrument, select TIM.
2. Press 1 (Fill cell).
3. Press the TIM980 filling cell button then press ✓.
Empty the Karl-Fischer cell (TIM980 only)Empty the Karl-Fischer cell (TIM980 only)
Empty the Karl-Fischer cell (TIM980 only)Empty the Karl-Fischer cell (TIM980 only)
Empty the Karl-Fischer cell (TIM980 only)
1. For Instrument, select TIM.
2. Press 2 (Empty cell).
3. Press the TIM980 emptying cell button then press ✓.