Programming guidelines 45
Edit method screenEdit method screen
Edit method screenEdit method screen
Edit method screen
1. Press 4 Method library.
2. Press , select the method to be edited from the list then press 2
Edit method.
3. Edit the necessary parameters.
a. Select the method mode. The modes available depend on the
Titration Manager used:
Measurement (pH, mV or mV at imposed current)
End point (titration to a preset mV or pH end point)
Coupled (methods chained in a same beaker)
Continuous IP (Inflection point detection with continuous addition
of the titrant)
Monotonic IP (Inflection point detection with incremental addition
of the titrant, increments of equal size)