52 Chapter 4
Run manual dosingRun manual dosing
Run manual dosingRun manual dosing
Run manual dosing
Refer to Short-Form Reminder no. 10.
1. Select the method.
2. Install the electrode group to be used to perform the measurement
(if measurements are to be done).
3. Install the reagent group to be used to perform the manual dosing.
4. Prepare the sample solution and place on the sample stand.
5. Display the Reagents window and press 5.
6. Select the reagent in the list of installed reagents and select the
electrode in the list of connected electrodes (if measurements are to
be done).
7. Enter an increment size, increment speed and maximum volume to
be dosed.
8. Select the type of measurement to be performed (pH, mV, mV (i>0)
or no measurement).
9. Press 1 to start.