Programming guidelines 35
7. Select the electrode Type, for electrodes with pH or mV functions:
Single pH - pH electrode (no reference part),
Combined pH - pH electrode with a reference part (with or
without temperature sensor),
Single metal/redox - mV (i=0) function electrode with a refer-
ence part,
Combined metal/redox - mV (i=0) function electrode with a
reference part (with or without temperature sensor),
Single ISE - mV (i=0) function ion selective electrode with no
reference part,
Combined ISE - mV (i=0) function ion selective electrode with a
reference part (with or without temperature sensor).
8. When a combined electrode has been defined, the instrument will
prompt you to specify if it has a built-in temperature sensor or not.
When a single electrode has been defined, the instrument will
prompt you to define a reference electrode in the Edit
electrode screen.
9. When a combined or a reference electrode has been defined,
enter the potential (in mV) of the reference element versus the
Standard Hydrogen Electrode (SHE).
This parameter enables to run calibration with automatic buffer
recognition when working with different reference electrode systems.
On the next page, see the table giving the potential at 25°C versus
the SHE of a few "reference elements/filling solution" couples.
10. When a combined pH or a single pH electrodes has been defined,
enter the internal pH of the electrode (pH int).
11. Press 1 to confirm.
12. Edit the electrode parameters (see page 37).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Radiometer Analytical TitraLab TIM960 and is the answer not in the manual?

Radiometer Analytical TitraLab TIM960 Specifications

General IconGeneral
BrandRadiometer Analytical
ModelTitraLab TIM960
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals