31
Cycle of automatic calibration proceeds as follows:
- balance software detects necessity of calibration and signalizes it by marker of
Celsius degree or by black marker
- from this moment time of 5 minutes is counted in which weighings can be
performed
- when this time elapses display indicates CAL_30 message and starts count
down 30..29..28 do 0 (indicated value is the counter)
- user has 30 seconds to make a decision:
if calibration is to be performed, does not take any actions
if one wants to finish measurements presses TARE key, after its pressing
balance returns to weighing showing previous result of weighing (possibility
of finishing series of measurements) in 5 minutes balance will indicate
CAL_30 message again
- process of calibration can be postponed frequently, but fact that long postponing
of calibration can be cause of bigger error during weighing should be noticed.
These errors are effect of changes of temperatures and as consequence
changes of balance sensitivity.
Automatic system includes 3 possibilities:
- calibration with regard to changes of temperature
balance is provided with precise system of controlling temperature
changes, temperature of calibration is registered every time, next is started
if temperature of surrounding changes 0,8
o
C
- calibration with regard to passage of time
user can declare 4 spaces of time that makes criterion for calibration,
following option are accessible: calibration every 1 - 12 hours.

Table of Contents

Other manuals for RADWAG AS series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS series and is the answer not in the manual?

RADWAG AS series Specifications

General IconGeneral
DisplayLCD with backlight
Maximum CapacityVaries by model
ReadabilityVaries by model
Pan SizeVaries by model
CalibrationInternal or external, depending on model
Stabilization Time≤ 3 seconds
InterfaceRS232, USB
Repeatability (Standard deviation)Varies by model
LinearityVaries by model
Net WeightVaries by model
Protection ClassIP 43

Related product manuals