11.4.2. External calibration
The external calibration should be performed with external mass class:
- E
2
– for balances AS
1. Start external calibration process
balances with internal calibration balances with external calibration
2. Order to load off the pan appears on the display (no load on the pan).
After yesing load off the pan press the ENTER key.
3. The balance determines mass of empty pan
4. Put load and press the ENTER
5. After the calibration the balance returns to submenu P1.2 uCAL
6. Return to weighing – as in the point 5.1.2.
If balance is verified user does not have the possibility to perform external
calibration process.
11.5. Calibration report printout
After calibration user can receive the calibration report. The report can be printed
on connected printer and sent to computer or recorded in file.
P1.6 CALr: no – report is not printed
P1.6 CALr: YES – report is printed
If the parameter has the value YES, the report is generated and sent
automatically.
A content of report depends on setting in submenu GLP. All options with YES
attribute are printed.

Table of Contents

Other manuals for RADWAG AS series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS series and is the answer not in the manual?

RADWAG AS series Specifications

General IconGeneral
DisplayLCD with backlight
Maximum CapacityVaries by model
ReadabilityVaries by model
Pan SizeVaries by model
CalibrationInternal or external, depending on model
Stabilization Time≤ 3 seconds
InterfaceRS232, USB
Repeatability (Standard deviation)Varies by model
LinearityVaries by model
Net WeightVaries by model
Protection ClassIP 43

Related product manuals