32
11.2. Entering calibration menu
P1 CAL
P1.1 iCAL |
P1.2 uCAL |
P1.3 tCAL |
P1.4 ACAL | both
P1.5 CALt | 1 h
P1.6 CALr | YES
P1.1 iCAL – Internal calibration
start of internal calibration process, process is completely automatic without
interference of balance operator, if pan is loaded display will indicate message
about necessity of removing the load.
P1.2 ECAL – External calibration
calibration with external weight, which value is saved in factory menu of
balance, function unavailable in verified balances.
balances with internal calibration balances with external calibration
P1.3 tCAL - Test of calibration
comparison of internal calibration mass with its value saved in balance memory.
balances with internal calibration balances with external calibration